Provided with
an Automatic Air Release Time
Setting Function
A top-open lid, which can be operated easily with a single hand or foot, is provided with the SX autoclaves. To open the lid, the lid lock is released by stepping on the foot pedal locates at the bottom while pressing the lid down. The lid can be opened with minimal effort owing to mechanical assistance. In addition, the lid can be closed easily just by pressing it down.
Once turned on the system automatically sets and displays a pre-set value. Operation is done simply by pressing the start switch. ES-315/ES-215 enables the set temperature and time for each mode program to be stored in its memory.
*The control panel lighting image shown above may vary from actual display.
Sterilization Mode
It contains a delay start timer function that allows operation to start 1 to 99 hours later.
Sterilization/warming Mode
This mode is useful when not removing culture media from the autoclave immediately after sterilization. Sterilized media can be kept warm to prevent coagulation.
Heating Mode
This mode program automates the dissolution of culture media that can sometimes be a painful process when using a water bath to boil them.
All models are supplied with stainless steel wire mesh baskets. No protrusions on inner surface of the chamber enable smooth loading and unloading of the basket as well as block design of all pipe fittings ensures safe operation.
The exhaust bottle to capture and store exhaust steam is placed on the front panel of the autoclave unit. Access to the chamber for replacing water inside, as well as for routine maintenance and inspection, is done on the front of the unit, which also permits operators to replace bellows, saving costs for services by the service personnel.
In case of abnormal chamber temperature rise, the autoclave stops its operation. In case of abnormal decrease in temperature, the sterilization timer is temporarily suspended and then reactivated when the set temperature is reached.
If overpressure in the chamber is detected, the autoclave stops all the function and indicates an error message simultaneously.
If the temperature inside the chamber exceeds the set temperature by 5°C or more, the autoclave stops all the function and indicates an error message simultaneously to avoid overheating.
The thermostat switch detects the temperature of chamber exterior wall, which allows the autoclave to stop all the function if the temperature exceeds 130°C.
Model | ES-215 | ES-315 | |
Operating temperature range (during sterilizing) |
105 - 123°C | 105 - 132°C | |
Operating pressure range | 0 - 127kPa | 0 - 186kPa | |
Maximum operating pressure | 147kPa | 216kPa | |
Temperature control | Digital, microprocessor controlled | ||
Temperature display / display range | Digital / -15 - 180°C | ||
Pressure display / display range | Analog / 0 - 250kPa | Analog / 0 - 400kPa | |
Heat source | 1.5kW electric heater | 2.0kW electric heater | |
Timer control | Digital, microprocessor controlled | ||
Timer display | Digital | ||
Timer display range |
Sterilizing mode: | 1 - 240min. or continuous | |
Sterilizing-warming mode: | 1 - 240min. (sterilizing) / 1 - 8hours (warming) | ||
Heating mode: | 1 - 8hours | ||
Start-timer setting range | 1 - 99hours in each operation mode | ||
Memory function | A set of temperature and time setting in each operation mode | ||
Safety devices | • Inside the chamber overheat prevention, • Outer wall of the chamber overheat prevention, • Overpressure prevention, • Temperature sensor disconnection prevention, • Empty heating prevention, • Leakage breaker, • Safety valve | ||
Malfunction prevention devices |
• Low water level detection, • Exhaust valve knob open/close detection, • Insufficient sterilization detection, • Chamber lid open/close detection |
||
Leakage breaker |
Rated breaking current: | 20A (120V) , 15A (220 / 230 / 240V) | 30A (120V) , 15A (220 / 230 / 240V) |
Rated sensed current: | 30mA (120V) , 10mA (220 / 230 / 240V) | ||
Protection type | Class 1 equipment | ||
Operating environment |
Ambient temperature: | 5 - 35°C | |
Relative humidity: | 30 - 85% | ||
Atmospheric pressure: | 700 - 1060hPa | ||
Gradient: | Within 2° | ||
Dimensions | φ248 x 543mm | φ325 x 740mm | |
Usable dimension | φ236 x 450mm | φ312 x 635mm | |
Capacity of chamber | 22ℓ | 53ℓ | |
Chamber material | SUS304 | ||
Capacity / material of exhasut bottle | 3ℓ / polyethylene | ||
Type / material of lid gasket | Accommodating internal pressure / silicone rubber | ||
Dimensions of main unit (mm) | 400W x 460D x 920H (Height from floor to control panel:705) | 490W x 560D x 1090H (Height from floor to control panel:875) | |
Net weight | 50kg | 80kg | |
Rated Voltage | AC 120 / 220 / 230 / 240V | ||
Power input | 15A (120V), 7A (220 / 230 / 240V) | 20A (120V), 10A (220V), 9A (230 / 240V) | |
Power supply requirements | Single-phase 120V AC (50 / 60Hz) 15A or above | Single-phase 120V AC (50 / 60Hz) 20A or above | |
Single-phase 220 / 230 / 240V AC (50 / 60Hz) 15A or above | |||
Power Comsumption (calorific value) | 1.5kW (1290kcal/h) | 2kW (1720kcal/h) | |
Accessories |
• Stainless steel basket with slatted bottom plate 1, • Water level sensor 1, • Inspection sheet 1, • Warranty card 1, • Customer card 1, • Operator's manual 1, • Clear plastic accessory case 1, • Attachment screw for clear plastic accessory case 1 |
||
Stainless basket 1 (φ224 x 376mm) | Stainless basket 2 (φ300 x 182mm) |
* For dimension details on the chamber and usable or inside dimension, please refer to each part dimensional drawing on the back page.