จรัญเอสโซซิเอทส์ (Charan Associates)

ตู้อบ -Clifton NE9-112S

฿1.00
ตู้อบ - Clifton NE9-112S
เหลือ 0 ชิ้น
ซื้อเลย
หยิบลงตะกร้า
  • หมวดหมู่ : ตู้อบลมร้อนในห้องแล้ป / Oven, Laboratory
  • รหัสสินค้า : 001224

รายละเอียดสินค้า ตู้อบ -Clifton NE9-112S

NE9-112S

 

112 Litre Drying Oven, Fan Assist Circulation.

« Return to Drying Ovens

 

Dimensions

External: 650w x 700d x 845h mm
Internal: 460w x 450d x 540h mm
Packed: 80x80x105cm
Capacity
Nominal Tank Capacity: 112 litre chamber Working capacity of the chamber can be smaller.
Temperature Range
ambient +5°C to 300°C
Sensitivity
±0.5°C @ 70°C / ±0.5°C @ 150°C Sensitivity measurements based on temperature readings on shelf in centre of the chamber.
Uniformity
±0.7°C @ 150°C Uniformity measured in vertical axis in the chamber using Wired Shelves.
Door configuration
Single door
Shelves fitted
2
Shelves max. no.
7
Power Consumption (220-240V)
2400W
Voltage
230V 50Hz Also available 230V 60Hz or 120V 60Hz models, please specify on order.
Packed Weight
85kg
  
Product Code
NE9-112S
Description
Clifton Drying Ovens feature intuitive easy to use controls with on screen guidance within a bold illuminated LCD display, offering precise time and temperature control from ambient +5°C to 300°C.

The Oven can run at a set temperature immediately, delay start or commence to start in the future at a given date and time. Additionally with three available programs with up to 6 segments in each program which allows the creation of temperature time profiles of your choice.

All control surfaces feature smooth easy to wipe clean surfaces.

The digital microprocessor controls feature real time clock, diagnostics, digital LCD display to 0.1°C resolution, dual pt100 probes, including event log and data log of actual temperatures along with USB and RS232 interface. Every unit is supplied with a Calibration Certificate for validation.

Featuring a stainless steel interior with stainless steel outer case powder painted. Particular attention paid providing a good tight door closure and hinge design to maintain a sealed chamber in use. Case is well insulated to reduce heat loss in use. Access port fitted as standard.

Controls
Featuring bold illuminated LCD graphic display of measured temperatures in real-time, in increments of ±0.1°C, with fan speed status, set temperature, date and time.

Intuitive to use controls, utilising illuminated touch sensitive keypads with display prompts assisting the setting of functions within the unit.

Option to set temperature in °C or °F.

The Oven can be set to run at a set temperature immediately, delay start [dd:hh:mm] or commence to start in the future at a given time and date [hh:mm, yyyy:mm:dd].

With three available programs and up to 6 segments in each it allows the creation of temperature time profiles of your choice, including for example fan speed, exhaust air flap setting, fast achieve set temperature and a setting allowing programs to be cycled. Any of these programs they can run immediately, delay start [dd:hh:mm ] or commence to start in the future at a given time and date [hh:mm, yyyy:mm:dd]. And when a program is running the controls are locked preventing accidental interruption or change however stopping a program is straightforward.

Heat-up Times
  • 15 mins
  • 20 mins
  • 30 mins
Fan speed
The Fan speed can be set from 0% no air convection within the chamber to 100% for maximum air convection.
Exhaust air flap
The Exhaust air flap can be set from 0% closed to 100% open and allows air from the chamber to vent.
Logger
Data logging of temperatures with date and times and an event log records door opening and closing, power on and off, starting, stopping and ending a program and alarms.
Safety Features
Switching on, the controls undergo an automatic diagnostic check.

Features a temperature sensor fail alarm with error reporting on LCD screen.

Power failure control with auto resume and event recording.

Setting of High and Low temperature Alarm, settable 0-5°C about set temperature.

Door opening generates an audible beep and indication on LCD display. When a program is running and should the door be left open a Door Alarm occurs, illuminating a red warning indicator and audible beeping until closed. Door open Alarm settable 0-5 minutes.

Over temperature protection can be selected in controls: Class 1 for overall unit protection or Class 2 for experiment/sample protection. Once protection is activated heating is switched off and a warning appears on screen. Class 3 protection, an accessory factory fit option, provides active protection once the temperature in chamber returns to normal values heating and control resumes.

Safety Class
Fitted with Class 1 for overall unit protection and Class 2 for experiment/sample protection. Accessory factory fit option Class 3 provides self-resetting protection at set temperature.