Multiparameter water quality meter contains 15 optional parameters. The meter is suitable for measuring the pH, ORP, Ion, Conductivity, TDS, Salinity, Resistivity and Dissolved Oxygen of liquids.
pH Mode:
• Multiparameter water quality meter is equipped with a large backlit LCD display.
• 1 to 5 points push-button calibration with auto-buffer recognition.
• Selectable pH buffer standard (USA/NIST/DIN) or using the custom calibration solutions.
• Automatic electrode slope display helps user decide whether to replace sensor.
ORP Mode:
• Single point offset calibration allows adjusting the displayed value to a known standard.
• Relative and absolute millivolts measurement modes provide accurate ORP readings.
Ion Mode:
• 2 to 5 points calibration including the eight concentration calibration points can be selected.
• Automatically recognize ion selective electrodes, does not need to specify type of ion.
• Direct ion concentration readout simplifies the elaborate measurement process.
• Selectable multiple concentration units include the ppm, mg/L and mol/L.
Conductivity/TDS/Salinity/Resistivity Mode:
• 1 to 5 points push-button calibration with automatic calibration solution recognition.
• Selectable cell constant (0.1/1/10), normalization temperature (20/25°C), TDS conversion
factor, seawater and practical salinity measurement modes.
• Linear or pure water compensation improves measurement accuracy of meter.
Dissolved Oxygen Mode:
• 1 or 2 points calibration using the air-saturated water and zero oxygen solution.
• Manual salinity and barometric pressure compensation improve the measurement accuracy.
Other Features:
• Automatic Temperature Compensation provides accurate reading over the entire range.
• Calibration Due Reminder prompts user to calibrate the meter regularly.
• Stability indicator automatically shows current measurement status.
• Auto-Hold function freezes stable measured value for easy viewing and recording.
• Manual temperature calibration corrects temperature deviation.
• Selectable temperature unit (°C or °F) meets different application requirements.
• Automatic electrode diagnosis shows zero point offset and slope of the electrode.
• Help message as a operational guide that helps you quickly begin using the meter.
• System menu allows setting the 15 parameters, including the number of calibration points,
resolution, stability condition, measurement unit, etc.
• Reset feature automatically resumes all parameters back to factory default settings.
• Expanded memory stores and recalls up to 500 readings.
• Built-in real-time clock stamps stored data to meet GLP standard.
• Stored data can be transferred into computer by USB communication interface.
Model |
Bante900 |
|
pH |
||
Range |
-2.000~20.000pH |
|
Accuracy |
±0.002pH |
|
Resolution |
0.1, 0.01, 0.001pH, Selectable |
|
Calibration Points |
1 to 5 points |
|
Calibration Solutions |
USA, NIST, DIN or User-defined |
|
Temperature Compensation |
0~100°C, 32~212°F |
|
Compensation Mode |
Manual or Automatic |
|
mV |
||
Range |
-1999.9~1999.9mV |
|
Accuracy |
±0.2mV |
|
Resolution |
0.1, 1mV, Selectable |
|
Calibration Points |
1 point (Only for Relative mV mode) |
|
Calibration Range |
±200mV |
|
Ion |
||
Range |
0.001~19999ppm, mg/L, mol/L (Refer to the range of ISE) |
|
Accuracy |
±0.5% F.S (Monovalent), ±1% F.S (Divalent) |
|
Resolution |
0.001, 0.01, 0.1, 1 |
|
Calibration Points |
2 to 5 points |
|
Calibration Solutions |
0.001, 0.01, 0.1, 1, 10, 100, 1000, 10000ppm, mg/L, mol/L |
|
Temperature Compensation |
0~100°C, 32~212°F |
|
Compensation Mode |
Manual or Automatic |
|
Conductivity |
||
Range |
0~20.00, 200.0, 2000µS/cm, 20.00, 200.0mS/cm |
|
Accuracy |
±0.5% F.S |
|
Calibration Points |
1 to 5 points |
|
Calibration Solutions |
10µS/cm, 84µS/cm,1413µS/cm,12.88mS/cm,111.8mS/cm |
|
Temperature Compensation |
0~100°C, 32~212°F, Manual or Automatic |
|
Temperature Coefficient |
0.0~10.0%/°C |
|
Compensation Modes |
Linear or Pure Water |
|
Cell Constant |
K=0.1, 1, 10 or User-defined |
|
Normalization Temperature |
20°C or 25°C |
|
TDS |
||
Range |
0~100ppt (Max. 200ppt, depending on factor setting) |
|
Accuracy |
±1% F.S |
|
TDS Factor |
0.1~1.0 (Default 0.5) |
|
Salinity |
||
Range |
0~10ppt (Max. 80ppt) |
|
Accuracy |
±1% F.S |
|
Measurement Modes |
Seawater or Practical Salinity |
|
Resistivity |
||
Range |
0~100MΩ |
|
Accuracy |
±1% F.S |
|
Resolution |
0.01, 0.1, 1 |
|
Dissolved Oxygen |
||
Range |
0.00~20.00mg/L (or ppm) |
|
Accuracy |
±0.2mg/L |
|
Resolution |
0.01mg/L |
|
Calibration Points |
1 or 2 points |
|
Temperature Compensation |
0~50°C, 32~122°F, Automatic |
|
Pressure Correction |
60.0~112.5kPa, 450~850mmHg |
|
Salinity Correction |
0~50g/L |
|
% Saturation of Oxygen |
||
Range |
0.0~200.0% |
|
Accuracy |
±2.0% |
|
Resolution |
0.1, 1% |
|
Temperature |
||
Range |
0~105°C, 32~221°F |
|
Accuracy |
±0.5°C, ±0.9°F |
|
Resolution |
0.1°C |
|
Calibration Points |
1 point |
|
Calibration Range |
Measured value ±10°C |
|
Others |
||
Hold Function |
Manual or Automatic |
|
Stability Conditions |
Low or High |
|
Calibration Due |
0 to 31 days |
|
Power Off |
Manual or Automatic (10, 20, 30 minutes) |
|
Reset Function |
Yes |
|
Memory |
Stores up to 500 data sets |
|
Output |
USB Communication Interface |
|
Connector |
BNC, 6-pin Mini Plug |
|
Display |
LCD (130×110mm) |
|
Power Requirements |
DC5V, using AC adapter, 220VAC/50Hz |
|
Dimensions |
210(L)×188(W)×60(H)mm |
|
Weight |
1.5kg |