Specifications | |
---|---|
Measuring Range (SI-units) |
RadEye G-10-Ex: From 0.5µSv/ to 100mSv/h RadEye GF-10-Ex: From 5µSv/h to 3Sv/h |
Measuring Range (USA) |
RadEye G-Ex: From 50 µR/h to 10 R/h RadEye GF-Ex: From 0.5 mR/h to 300 R/h |
Sensitivity (Cs-137, 660 keV), (SI-units) |
RadEye G-10-Ex: ~1.7 cps/µSv/h RadEye GF-10-Ex: ~0.13 cps/µSv/h |
Sensitivity (Cs-137, 660 keV), (USA) |
RadEye G-Ex: ~17 cps/mR/h RadEye GF-Ex: ~1.3 cps/mR/h |
Dose | 0.5µSv to 10Sv [50 µR to 1000 R] |
Linearity Error | max. ± 10% in the measuring range |
Alarm Thresholds | Two each thresholds for dose and dose rate |
Energy Range | 48keV to 3MeV according to IEC60846-1 and ANSI 42.17A accordingly |
Working Temperature | -20° to +50°C |
Relative Humidity | 10 to 90% at 35°C |
Protection Degree | IP 65 according to EN 60 529 |
ATEX Classification | II 2G Ex ia IIB T4 IBExU10ATEX1096 |
Size | L x W x H: 96 x 61 x 31mm without rubber protector |
Weight | approx. 160g, including two batteries |
Memory | The latest 1600 measured values are saved and can be read out via PC-program. Logbook with 250 entries for changes of configuration, occurring alarms and errors. |
Cat. No. | Description |
---|---|
425067660 | RadEye G-10-Ex |
425067460 | RadEye G-Ex |
425067670 | RadEye GF-10-Ex |
425067470 | Radeye GF-Ex |