จรัญเอสโซซิเอทส์ (Charan Associates)

เครื่องวิเคราะห์ธาตุด้วยเปลวไฟ

฿0.00
เครื่องวิเคราะห์ธาตุด้วยเปลวไฟ ยี่ห้อ BWB รุ่น Flash
เหลือ 0 ชิ้น
  • หมวดหมู่ : เครื่องวิเคราะห์ธาตุด้วยเปลวไฟ / Flame photometer
  • รหัสสินค้า : 001734

รายละเอียดสินค้า เครื่องวิเคราะห์ธาตุด้วยเปลวไฟ

BWB Flash flamephotometer 


The BWB Flash Flame Photometer

The newest feature to Flame Photometry

Finally, a Flame Photometer specific to your needs. The first simultaneous Flame Photometer with built in compressor and configurable element analysis to your requirements.

 

The BWB Flash is the first and only fully configurable up to four channel flame photometer that detects and displays all calibrated ions at the same time.


The BWB Flash was created by combining the award-winning features of our XP model with new, innovative, and cutting-edge technology to provide unrivalled levels of accuracy, usability, and reliability while significantly reducing analysis time. When a different measurement is required, our customers do not need to waste valuable time changing filters and re-calibrating, as is a common requirement in all other Flame Photometer designs from 1951 to the present.


The Flash looks exceptionally modern, with clean lines and contemporary bright colour schemes that will fit into any laboratory. The instrument has a large 5” touch screen display that can be easily viewed from any angle and operated with most laboratory disposable gloves.


Our metal casework creates a sturdy portable instrument with nearly limitless future recycling possibilities. Of course, ROHS-compliant components and materials are used throughout.


The BWB Carbon Footprint is exceptionally low, in keeping with the company’s goals of being a responsible manufacturer, and we’ve succeeded in lowering the gas consumption rate by 13% while also lowering the electricity consumption (kWh) by 26% compared to previous model generations.


Unlike other manufacturers who charge extra for essential accessories, BWB includes everything needed to use the product right away. Simply add Gas!! (Propane, LPG, or butane) We also include data sharing via USB from the BWB Flash to your PC as standard. This connects the BWB Flame Emission Spectrometer (FES) output to the Internet or your company intranet systems, allowing for Distributed Control, which is critical for multinational corporations, universities, and government agencies.                                     


                 




Technical DataSample rate 

2.5 – 4ml/min

​

Tubing materials 

Silicone and Tygon®  

​

Required desk space for Instrument  

101*cm (H) x 45cm (W) x 35cm (D)

*For Safety reasons the Flame photometer requires 1m of unobstructed space above to allow dissipation of heat from the chimney.

​

Instrument size 

Flash Photometer
51cm (H) x 33cm (W) x 35cm (D)
(20in x 13in x 14in)

​

Shipping
65cm (H) x 49cm (W) x 43cm (D)
(25in x 19in x 17in)

​

Weight 

Flash Photometer – 10kg (22.1lbs)

Shipping – 18kg (39.7 lbs)*

*subject to vary based on accessories included

​

Optimal range

Single point calibration

Na 0.1 – 60ppm 

K 0.05 – 100ppm 

Li 0.05 – 50ppm 

Ca 1.0 – 100ppm

​

Multi Point Calibration (with dilution)

Na 0.1 – 1000ppm 

K 0.05 – 1000ppm 

Li 0.05 – 1000ppm 

Ca 1.0 – 1000ppm  

​

Reproducibility 

<1% Coefficient of variability for 20 consecutive samples over ten minutes at concentrations of 100ppm or less. (After instrument stabilisation). 


Specificity 

Na/K/Li = <1% to each other when equal in concentration at <100ppm

​

Limit of detection (LOD) and
limit of quantification (LOQ) 

LOD                     LOQ        

Na – 0.03ppm    Na – 0.1ppm

K – 0.02ppm      K – 0.05ppm

Li – 0.02ppm     Li – 0.05ppm

Ca – 0.3ppm      Ca – 1ppm

​

Time to stability 

Less than 15 seconds after sample is introduced into the flame.

​

Ions measured  

Simultaneous measurement of model specific elements

​

Interfaces  

USB

0-2.5 volt output (based on sample concentration linked to element of users choice)

Optional 4-20mA output in place of the above

Optional integrated printer

.csv and .pdf generated reports and files via FP-PC software

​

Recommended minimum warm up time based on ambient temperature  

21ºC – 25 minutes     

​

Power requirements   

100V – 250V AC at 50 or 60Hz automatically selecting 

​

Fuel requirement    

Propane, Butane or Natural Gas* regulated to 19Bar. Flow rate of 0.4l/min

*with modifications. BWB Technologies recommends either Propane or Butane for optimum results.

​

Readout  

5" colour capacitive touch screen display